Effect of Intermediate Agents and Pre-heating of Repairing Resin on Composite-repair Bonds

,
,
,
,
,
,
, and
Page Range: 363 – 371
DOI: 10.2341/06-105
Save
Download PDF

SUMMARY

This study investigated the composite-to-composite microtensile bond strength and interfacial quality after using different combinations of intermediate agents and pre-curing temperatures of repairing resin. Forty-five composite discs (8×4mm) of Gradia Direct Anterior (GC Corp), stored in a saline solution at 37°C for one month, were sandblasted (50 μm aluminum oxide), cleaned (35% phosphoric acid) and randomly divided into three groups (n=15) according to the intermediate agent applied: 1) no treatment; 2) unfilled resin (Scotchbond Multi-Purpose Adhesive, 3M ESPE); 3) flowable composite (Gradia LoFlo, GC Corp). Each disc was incrementally repaired (8×8 mm) with the same resin as the substrate. For each group, three subgroups (n=5) were created, depending on the pre-curing temperature of the repairing resin—4°C, 23°C or 37°C. Two bonded specimens per group were prepared to evaluate the composite-to-composite interfacial quality via scanning electron microscope. Microtensile bond strength measurements were performed with the remaining three specimens and failure mode was examined by stereomicroscopy. Two-way ANOVA revealed that temperature (p<0.001), intermediate agent (p<0.001) and the interaction (p=0.002) significantly affected the repair strength. Post-hoc comparisons revealed that bond strengths were significantly lower using a 4°C repairing resin in groups where intermediate agents were used. The highest bond strengths were recorded when flowable composite was used as an intermediate agent under each of the three temperature conditions. Interfacial quality improved by raising the resin temperature from 4°C to 37°C.

INTRODUCTION

In the last decade, growing demands by patients for mercury-free, esthetic restorations has markedly increased the use of direct, light-activated resin composites in restorative dentistry.1 However, despite the continuing development of resin composites with improved properties, several factors, such as discoloration, color mismatch, wear, chipping or bulk fracture may still present concerns.2–4 As a result, when esthetics or function are compromised, an operative treatment is needed, and the clinician must decide whether to replace or simply repair these restorations. When failure is limited to resin composite bulk and the tooth-resin interface is free of staining or secondary caries, replacement of the whole restoration would unnecessarily involve more complex, costly treatments, both in economic and biological terms.2 Minimally invasive approaches, such as selective composite repair or refurbishing, should be the preferred method.5

In order to achieve long-term clinical success, an intimate resin-to-resin coupling6 and high interfacial strength are required when repairing a composite restoration. Due to the limited amount of residual free-radicals available for reacting with new resin monomers on the old, contaminated, highly cross-linked resin matrix of composite substrate,7–9 chemical and/or mechanical treatments have been investigated for preparation of the surface to be repaired.10–15 Aluminum oxide sandblasting, followed by application of a bonding agent, proved to be a reliable method for enhancing composite repair strength.81215 Benefits in resin-to-resin bond seem to arise from the physico-chemical changes induced by the air-abrasion process. This results in a roughened composite surface, thus improving the surface area available for bonding.68 In addition, an increase in surface energy is claimed to facilitate wettability of the intermediate agent,16 thus promoting a closer adaptation to the composite substrate.81013 As a consequence of the penetration of resin monomers into surface irregularities, a micromechanical interlocking is achieved.

Filled 817–18 and unfilled 69–1319–21 resins usually have been applied as intermediate agents for composite repair, thus achieving comparable results in terms of interfacial quality and strength.17–18 Recently, the application of flowable composites as intermediate agents with19–20 or without18 a preliminary adhesive coating was demonstrated to be a valid alternative to traditional bonding procedures for composite repair.

Resin composites are traditionally stored at temperatures ranging from 2°C to 25°C, as suggested by manufacturers, and retrieved from the storage place immediately prior to use. As resin viscosity is temperature-dependent and directly related to the handling characteristics of the composite,22–24 differences in ease of placement and shaping on the repaired site could be expected in a clinical situation, potentially affecting interfacial quality and repair strength.

This study evaluated the effect of different pre-curing temperatures of repairing resin on composite-to-composite bond strength and interfacial micromorphology. The tested null hypothesis is that different combinations of intermediate agents and storage temperatures of the repairing resin do not affect the adhesion between an aged composite substrate and a fresh composite material.

METHODS AND MATERIALS

Forty-five cylinder-shaped composite discs were made by layering 1 mm thick increments of a micro-filled hybrid resin composite (Gradia Direct Anterior, GC Corp, Tokyo, Japan, shade A3) in a split aluminum mold (8 mm deep × 8 mm in diameter). An internal piston was adjusted to a depth of 4 mm in order to standardize the height of the composite discs. Each increment was condensed with a clean plastic filling instrument to avoid contamination and light-cured for 20 seconds (VIP, BISCO Inc, Schaumburg, IL, USA, output: 600 mW/cm2). The last increment was covered with a Mylar strip and compressed with a glass microscope slide in order to obtain a flat surface. Each specimen was removed from the mold, additionally light-cured on the portions previously in contact with the metallic surface of the mold and stored in saline solution at 37°C for one month. One flat composite surface of the cylinder was selected, then sandblasted with 50 μm aluminum oxide powder for 10 seconds at a working distance of 5 mm at a pressure of 5.5·105 Pascals (Pa) (Microetcher II, Danville Engineering Inc, San Ramon, CA, USA). For purposes of cleaning, a 35% phosphoric acid etchant (Scotchbond etchant, 3M ESPE, St Paul, MN, USA) was applied for 30 seconds. After water rinsing and air drying, composite discs were randomly divided into three experimental groups (n=15) according to the different intermediate agents used before applying the repairing resin: Group 1: direct bonding between old and fresh composite; Group 2: application of an unfilled bonding agent (Scotchbond Multi-Purpose Adhesive, 3M ESPE); Group 3: application of a low-viscosity flowable composite (Gradia LoFlo, GC Corp). The tested intermediate agents were stored and used at room temperature (23°C).

For each experimental group, mono-dose tips containing the uncured repairing resin (Gradia Direct Anterior) were stored at different temperatures before use: 4°C (refrigerator); 23°C (room temperature); 37°C (incubator), leading to three subgroups (n=5).

Each specimen was returned to the mold, and a new restoration was packed onto the previously treated surface by layering 1 mm thick increments of the repairing resin and obtaining an 8 mm high cylindrical repaired sample. The repairing material was the same as the respective composite substrate except for its shade (Gradia Direct Anterior, shade B3). This was intentionally done to facilitate the positioning of the repair interface of microtensile specimens in the testing machine and to visually identify the failure site, particularly when no intermediate agent was applied.

All the materials were used according to manufacturers' instructions. Batch numbers, application mode and chemical composition are reported in Table 1.

Table 1 Batch Numbers, Composition and Application Mode of the Materials Tested in the Study
Table 1

Composite Repair Microtensile Testing

After a 24 hour-storage in saline solution at 37°C, three bonded specimens from each subgroup were randomly selected and cut with a slow-speed diamond saw under water cooling (Isomet, Buehler, Lake Bluff, IL, USA) to produce multiple beam-shaped sticks with a cross-sectional area ranging between 0.79 mm2 and 0.81 mm2 according to the non-trimming technique of the microtensile bond strength test.25 The sticks consisted of the old composite in one-half of their length and the new one in the remaining half. Each beam was glued at the ends of a Gerardeli's jig26 with cyanoacrylate glue (Zapit, Dental Ventures of America, Inc, Corona, CA, USA) and loaded in tension at a crosshead speed of 0.5 mm/minute in a universal testing machine (Controls, Milano, Italy). The load at failure was recorded in Newtons (N). The cross-sectional area at the site of fracture was measured to the nearest 0.01 mm with a digital caliper in order to express bond strength values in MegaPascals (MPa). The fractured beams were evaluated under a stereomicroscope (SMZ645, Nikon Co, Tokyo, Japan) at 50× magnification to record failure modes, classified as cohesive (within the old/repairing resin composite), adhesive (at intermediate agent/resin composite interface) or mixed (a combination of cohesive/adhesive failure modes). When no intermediate agent was applied, adhesive failure was considered to be the fracture occurring at the composite-to-composite interface. The number of sticks representative of the three failure patterns within each experimental subgroup was expressed as a percentage.

Statistical Analysis

All the sticks that failed prematurely (two sticks for no intermediate agent/4°C subgroup and one stick for Scotchbond Multi-Purpose/4°C subgroup) were considered as “zero bond” values and were included in the statistical calculations. After analyzing the bond strength data for normality of data distribution (Kolmogorov-Smirnov test) and homogeneity of variances (Levene's test), a two-way ANOVA was applied, with bond strength as the dependent variable, intermediate agent and repairing resin temperature as factors. The Tukey test was used for post-hoc multiple comparisons. All analyses were processed using SPSS 11.0 software (SPSS, Chicago, IL, USA). The level of significance was set at p<0.05.

Scanning Electron Microscopy (SEM) Evaluation

The two remaining bonded specimens in each subgroup were processed for composite-to-composite interfacial characterization. Each specimen was cut perpendicular to the bonded interface with a slow-speed saw under water cooling (Isomet, Buehler), resulting in five 1 mm thick slabs. Each slab was then polished with SiC papers of increasing grit (#600, #1000, #1200, Buehler), rinsed with deionized water, air-dried, mounted on aluminum stubs and sputter-coated with gold (Polaron Range SC7620, Quorum Technology, Newhaven, England). The samples were observed under a scanning electron microscope (JSM 6060 LV, JEOL, Tokyo, Japan) at different magnifications.

RESULTS

Composite-to-Composite Microten-sile Bond Strength

The results of microtensile bond strength testing for each experimental group are summarized in Table 2. Two-way ANOVA revealed that the temperature (p<0.001), the intermediate agent (p<0.001) and the interaction between these two factors (p=0.002) significantly affected composite repair strength. Bond strengths were significantly lower, using a 4°C-repairing resin in groups where intermediate agents were not used. The highest results were recorded when flowable composite was used as an intermediate agent under each of the three temperature conditions of the repairing resin. In groups where no intermediate agent was used, bond strengths were the lowest when using a 4°C- and a 23°C-repairing resin. However, when Gradia Direct Anterior was preheated at 37°C, statistically superior values were achieved, which were comparable to bond strengths in adhesive-mediated groups.

Table 2 Microtensile Bond Strength of the Tested Groups
Table 2

Stereomicroscopy and SEM Analysis

The distribution of failure modes in the tested groups is reported in Figure 1. Interfacial failure at the composite-to-composite interface mainly occurred when no intermediate agent was applied, regardless of the precuring temperature of the repairing material. Cohesive and mixed failures were mostly recorded when Scotchbond Multi-Purpose was used. Adhesive and mixed failures were predominant in the Gradia Direct LoFlo/4°C subgroup, while a frank cohesive failure in substrate/repairing resin was the main failure mode when Gradia Direct LoFlo was covered by 23°C- and 37°C-composite. No failures occurred within the flowable resin.

Figure 1. Distribution of failure modes for all experimental groups. Values are the percentage of sticks fractured in the respective failure mode within each subgroup.Figure 1. Distribution of failure modes for all experimental groups. Values are the percentage of sticks fractured in the respective failure mode within each subgroup.Figure 1. Distribution of failure modes for all experimental groups. Values are the percentage of sticks fractured in the respective failure mode within each subgroup.
Figure 1 Distribution of failure modes for all experimental groups. Values are the percentage of sticks fractured in the respective failure mode within each subgroup.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

When direct composite-to-composite coupling was performed, SEM evaluation of bonded interfaces showed an effect of resin temperature on interfacial quality. Gaps and defects decreased by increasing the repairing resin temperature from 4°C to 37°C (Figure 2A,B). When Scotchbond Multi-Purpose was used as the intermediate agent, interfacial defects visible at the adhesive/4°C-composite interface (Figure 3A) disappeared by layering 23°C- and 37° C- composite, showing a better interfacial adaptation (Figure 3B). While layering 4°C-cooled composite over flow-able Gradia LoFlo resulted in imperfect coupling (Figure 4A), the use of 23°C- and 37°C-composite allowed for a more uniform, complete composite-to-composite adaptation (Figure 4B).

Figure 2. Representative Gradia-to-Gradia interfaces repaired by direct bonding of the repairing composite (RC) to the sandblasted composite substrate (CS) (bar=50 μm). Figure 2A. SEM micrograph of CS repaired by using 4°C-cooled RC. Gaps (20–100 μm) and discontinuities interrupt the interfacial coupling. Figure 2B. SEM micrograph of CS repaired with 37°C pre-heated RC. A good interfacial composite-to-composite adaptation barely interrupted by small defect (less than 15 μm) is detectable.Figure 2. Representative Gradia-to-Gradia interfaces repaired by direct bonding of the repairing composite (RC) to the sandblasted composite substrate (CS) (bar=50 μm). Figure 2A. SEM micrograph of CS repaired by using 4°C-cooled RC. Gaps (20–100 μm) and discontinuities interrupt the interfacial coupling. Figure 2B. SEM micrograph of CS repaired with 37°C pre-heated RC. A good interfacial composite-to-composite adaptation barely interrupted by small defect (less than 15 μm) is detectable.Figure 2. Representative Gradia-to-Gradia interfaces repaired by direct bonding of the repairing composite (RC) to the sandblasted composite substrate (CS) (bar=50 μm). Figure 2A. SEM micrograph of CS repaired by using 4°C-cooled RC. Gaps (20–100 μm) and discontinuities interrupt the interfacial coupling. Figure 2B. SEM micrograph of CS repaired with 37°C pre-heated RC. A good interfacial composite-to-composite adaptation barely interrupted by small defect (less than 15 μm) is detectable.
Figure 2 Representative Gradia-to-Gradia interfaces repaired by direct bonding of the repairing composite (RC) to the sandblasted composite substrate (CS) (bar=50 μm). Figure 2A. SEM micrograph of CS repaired by using 4°C-cooled RC. Gaps (20–100 μm) and discontinuities interrupt the interfacial coupling. Figure 2B. SEM micrograph of CS repaired with 37°C pre-heated RC. A good interfacial composite-to-composite adaptation barely interrupted by small defect (less than 15 μm) is detectable.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

Figure 2. Continued.Figure 2. Continued.Figure 2. Continued.
Figure 2 Continued.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

Figure 3. Representative Gradia-to-Gradia interfaces repaired by unfilled resin Scotchbond Multi-Purpose Adhesive as an intermediate agent (RC: repairing composite adhesive resin, CS: composite substrate) (bar=50 μm). Figure 3A: Voids and defects (10–70 μm) both in the RC and in close proximity of the adhesive layer (A) are detectable with 4°C-RC. Figure 3B. A uniform adaptation at the CS-A and A-RC interfaces are evident with 23°C-RC.Figure 3. Representative Gradia-to-Gradia interfaces repaired by unfilled resin Scotchbond Multi-Purpose Adhesive as an intermediate agent (RC: repairing composite adhesive resin, CS: composite substrate) (bar=50 μm). Figure 3A: Voids and defects (10–70 μm) both in the RC and in close proximity of the adhesive layer (A) are detectable with 4°C-RC. Figure 3B. A uniform adaptation at the CS-A and A-RC interfaces are evident with 23°C-RC.Figure 3. Representative Gradia-to-Gradia interfaces repaired by unfilled resin Scotchbond Multi-Purpose Adhesive as an intermediate agent (RC: repairing composite adhesive resin, CS: composite substrate) (bar=50 μm). Figure 3A: Voids and defects (10–70 μm) both in the RC and in close proximity of the adhesive layer (A) are detectable with 4°C-RC. Figure 3B. A uniform adaptation at the CS-A and A-RC interfaces are evident with 23°C-RC.
Figure 3 Representative Gradia-to-Gradia interfaces repaired by unfilled resin Scotchbond Multi-Purpose Adhesive as an intermediate agent (RC: repairing composite adhesive resin, CS: composite substrate) (bar=50 μm). Figure 3A: Voids and defects (10–70 μm) both in the RC and in close proximity of the adhesive layer (A) are detectable with 4°C-RC. Figure 3B. A uniform adaptation at the CS-A and A-RC interfaces are evident with 23°C-RC.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

Figure 3. Continued.Figure 3. Continued.Figure 3. Continued.
Figure 3 Continued.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

Figure 4. Representative Gradia-to-Gradia interfaces repaired by flowable composite Gradia LoFlo as an intermediate agent (RC: repairing composite, FC: flowable composite, CS: composite substrate) (bar=50 μm). Fiigure 4A. Particular of 4°C RC-FC interface. A gap (50–100 μm), probably resulting from difficult adaptation of RC is clearly shown. Figure 4B. 23°C-composite as RC. Good coupling at both CS-FC and FC-RC interfaces is visible.Figure 4. Representative Gradia-to-Gradia interfaces repaired by flowable composite Gradia LoFlo as an intermediate agent (RC: repairing composite, FC: flowable composite, CS: composite substrate) (bar=50 μm). Fiigure 4A. Particular of 4°C RC-FC interface. A gap (50–100 μm), probably resulting from difficult adaptation of RC is clearly shown. Figure 4B. 23°C-composite as RC. Good coupling at both CS-FC and FC-RC interfaces is visible.Figure 4. Representative Gradia-to-Gradia interfaces repaired by flowable composite Gradia LoFlo as an intermediate agent (RC: repairing composite, FC: flowable composite, CS: composite substrate) (bar=50 μm). Fiigure 4A. Particular of 4°C RC-FC interface. A gap (50–100 μm), probably resulting from difficult adaptation of RC is clearly shown. Figure 4B. 23°C-composite as RC. Good coupling at both CS-FC and FC-RC interfaces is visible.
Figure 4 Representative Gradia-to-Gradia interfaces repaired by flowable composite Gradia LoFlo as an intermediate agent (RC: repairing composite, FC: flowable composite, CS: composite substrate) (bar=50 μm). Fiigure 4A. Particular of 4°C RC-FC interface. A gap (50–100 μm), probably resulting from difficult adaptation of RC is clearly shown. Figure 4B. 23°C-composite as RC. Good coupling at both CS-FC and FC-RC interfaces is visible.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

Figure 4. Continued.Figure 4. Continued.Figure 4. Continued.
Figure 4 Continued.

Citation: Operative Dentistry 32, 4; 10.2341/06-105

DISCUSSION

Significant differences in repair strength were found between the experimental groups, depending on the intermediate agent and composite temperature preceding photo-polymerization. When groups in which the repairing resin temperature was the same were compared, the highest composite-to-composite strengths were achieved using a flowable composite as an intermediate agent under each of the three temperature conditions. As the interaction between intermediate agent and temperature was significant, the effect of the repairing resin temperature depended on the intermediate agent applied and vice versa. Hence, the null hypothesis tested has to be rejected.

Rheology studies indicated that resin composites, being visco-elastic materials by nature, respond to an increase in external temperature according to the Arrhenius equation, which results in the exponential decrease in their viscosity.2427–28 A decline in viscosity of the repairing resin, achieved by raising the temperature from 4°C to 37°C, resulted in a critical factor in improving composite-to-composite interfacial coupling and strength.

A composite repair, based on direct bonding between the repairing resin and the sandblasted substrate, is not usually recommended. There are two factors that may impair adhesion: the low chemical bonding potential of the aged substrate7–9 and the incomplete penetration of the highly viscous fresh composite into surface pits and depressions.1321 Thus, the application of a bonding resin as intermediate agent is advisable for enhancing substrate wetting.1012 In this study, similar repair strength values were obtained by using cooled and room-temperature-stored composite as repair materials without applying any intermediate agent. However, in spite of comparable bond strength results, SEM analysis demonstrated an inferior composite-to-composite adaptation with 4°C-composite. Clear discontinuities could be more frequently observed compared to interfaces formed using 23°C-composite.

Conversely, superior repair strengths and a complete, more uniform coupling with the substrate, were achieved by warming the repairing resin composite to 37°C. The increased temperature from 23°C to 37°C presumably allowed the repairing composite to achieve optimal viscosity and flowing properties, thus promoting a deeper resin interdiffusion into surface microretentions. A significant improvement in repair strength was also achieved. The prevalence of adhesive failures with 4°C-composite (95%) and their progressive reduction with 23°C- (75%) and 37°C-composite (55%) seemed to confirm the beneficial contribution of pre-heated, low-viscosity repairing material to micromechanical retention. It is interesting to note that bond strengths achieved by direct bonding of the pre-heated resin were similar to those attained by applying the unfilled resin Scotchbond Multi-Purpose Adhesive as intermediate agent, regardless of the pre-curing temperature. However, the smaller percentage of adhesive failures recorded with Scotchbond Multi-Purpose (45% for 4°C-, 30% for 23°C- and 25% for 37°C-composite) may reveal the formation of a bonded interface that is more prone to withstand tensile stresses. Moreover, since composite-to-composite bond strength was higher when flowable resin was used at 23°C, composite repair without the use of an intermediate agent should not be routinely recommended, even if the repairing composite is warmed at 37°C.

A noteworthy finding of this investigation was the statistically superior performance of Gradia LoFlo as an intermediate agent. Since flowable composites possess lower mechanical and physical properties than traditional hybrid materials,29–31 their application in occlusal, high stress-bearing areas could represent a concern. Therefore, the use of flowables as intermediate agents prior to layering a repairing hybrid composite has been proposed.18–20 This allows flowable composites to overcome their potential clinical limitations, taking advantage of their stress-relieving behavior.32–33 Whereas adhesive coating preceded the application of flowable composite as an intermediate agent in some studies,19–20 a flowable resin was used as the sole interfacial material in a recent investigation.18 It may be argued that the high polymerization shrinkage of flowable composites could generate high interfacial stress,34 possibly impairing the composite-to-composite bond. However, no gaps or discontinuities were observed at the composite substrate-flowable intermediate interface. This may depend on the low intrinsic rigidity traditionally exhibited by flowable materials when compared to unfilled resins.34 As a consequence of their counteracting role, the interfacial integrity may be preserved.32–33 The higher thickness of the flow-able intermediate agent (about 200 μm) to that of the unfilled resin (about 25 μm) is an additional factor that may have improved the stress-relieving potential of the composite-repair joint.33–35 The predominance of cohesive failures within the substrate/repairing material, instead of adhesive failures, further confirms the good coupling achieved by applying flowable composite as a filled intermediate agent. Cohesive failures within the flowable resin were not recorded. This may be explained by the elasticity properties of the flow-able intermediate, allowing for a stress-bearing ability higher than that of the hybrid resin composite.29–303436 The intimate adaptation of Gradia LoFlo to the resin substrate is in agreement with the satisfactory results of marginal/internal adaptation achieved on enamel and dentin, using flowable composite as an intermediate layer in Class I, II and V cavities.37–39

The quality of the repaired interface was also influenced by the temperature of the repairing composite. Gaps, detected when the cooled composite was used, are likely to behave as short- and/or long-term stress-raisers, probably leading to repair failure. On the contrary, the application of an uncured, pre-heated composite over a layer of previously cured flowable resin was the optimal combination in terms of repair strength and interfacial quality.

It is known that the photopolymerization of resin-based materials, relying on a complex free-radicals formation process and leading to a highly cross-linked network formation, may be influenced by the precuring temperature. An increased temperature may enhance the radicals mobility and polymerization rate, resulting in a superior crosslinking density and final degree of conversion of dimethacrylate-based monomers.2240–41 As a result, a warmed composite is prone to exhibit greater polymerization shrinkage than a cooler material.32–3342 In this investigation, since the fabricated composite build-ups represented a favorable configuration factor (C-factor), with a low ratio between bonded and unbonded composite surfaces, resin shrinkage presumably did not play a major role in the recorded bond strengths.19 Hence, these results are clinically related to situations of low C-factor composite repair, such as the esthetic veneering of a pre-existing composite restoration because of color mismatch or unsatisfactory anatomic form. Clinical conditions with higher C-factor, such as the repair of restoration margins, could represent a concern, and this topic will be addressed in future investigations. The authors are aware that the amount of repairing composite in the current experimental setup might be viewed as a potential limitation, since the amount of resin placed in common clinical situations is smaller. Nevertheless, in the majority of methods used for bond strength measurement, specimen preparation procedures inevitably differ from clinical procedures. Comparably, these methods allow for the production of standardized test specimens and assess the potential differences between the investigated bonding approaches.25

On the other hand, bonding procedures performed on dental substrates, composite repair requires complete isolation of the operative field by a rubber dam. It has been reported that, by using the rubber dam, the intra-oral temperature around the treatment area becomes close to room temperature.43 In a clinical scenario, after composite pre-heating, a time delay exists between dispensing the material, its placement and shaping of the substrate and light-curing, possibly decreasing resin temperature.2242 Thus, the detrimental effect of polymerization shrinkage expected at 37°C could also be minimized in high C-factor clinical conditions by this gradual cooling process.40–41 However, even if pre-heated composite cools to 23°C before light-activation, benefits, in terms of interfacial strength and quality, are still present.

In this study, the effect of the pre-curing temperature was investigated only for a single brand of composite. As temperature may differentially affect the viscosity, depending on the resin composition (filler content, resin monomers), additional research should be done to determine if these results can be confirmed for other commercial materials.222428 Recently, a chair-side preheating device that allows the operator to warm resin composites at 50°C-60°C immediately before placement into a tooth cavity has been marketed.44 The effect of the additional heating on composite-to-composite strength and micromorphology should be further investigated.

Also, the clinician may not be aware of the type of resin composite needing re-intervention (especially when dealing with an old restoration performed by a different operator), and he or she may be faced with the question of materials compatibility. In this study, the composite substrate/repair and the flowable resin were based on UDMA monomers, while the bonding agent was based on Bis-GMA monomers. Although previous investigations have shown that a similar chemical nature of the bonded materials is not a prerequisite for superior interfacial coupling,45–46 the in vitro performance of homogeneous (similar composition) and heterogeneous (dissimilar composition) repairs mediated by bonding and flowable resins, based on different monomers, deserves major attention in future research.

CONCLUSIONS

Within the limitations of this in vitro study, the following conclusions can be drawn:

  1. The use of cooled (4°C) microfilled hybrid composite (Gradia Direct Anterior) as a repairing material resulted in poor composite-to-composite interfacial adaptation and lower repair strengths. The difficult adaptation of a highly viscous material to the roughened or intermediate agent-covered substrate probably explains these results.

  2. Pre-heating of the repairing material to 37°C significantly enhanced repair strengths when no intermediate agent was used. Lower viscosity, achieved by heating the composite, may also justify the uniform composite-to-composite interfacial adaptation observed.

  3. The use of a low-viscosity flowable composite (Gradia LoFlo) as an intermediate agent resulted in a significant improvement in repair strength under each of the three temperature conditions of the repairing resin.

Acknowledgments

The authors thank GC Corp and 3M ESPE for supplying the materials used in this study.

References

  • 1
    Roulet, J. F.
    and
    M.Degrange
    . 2000. Chapter 15–Composite resin restorations on posterior teeth.in. Adhesion: The Silent Revolution in Dentistry.
    Chicago
    Quintessence
    . p. 253.
  • 2
    Mjör, I. A.
    and
    V. V.Gordan
    . 2002. Failure, repair, refurbishing and longevity of restorations.Operative Dentistry27
    5
    :528534.
  • 3
    Vichi, A.
    ,
    M.Ferrari
    , and
    C. L.Davidson
    . 2004. Color and opacity variations in three different resin-based composite products after water aging.Dental Materials20
    6
    :530534.
  • 4
    Kolbeck, C.
    ,
    M.Rosentritt
    ,
    R.Lang
    , and
    G.Handel
    . 2006. Discoloration of facing and restorative composites by UV-irradiation and staining food.Dental Materials22
    1
    :6368.
  • 5
    Tyas, M. J.
    ,
    K. J.Anusavice
    ,
    J. E.Frencken
    , and
    G. J.Mount
    . 2000. Minimal intervention dentistry–a review FDI Commission Project 1–97.International Dental Journal50
    1
    :112.
  • 6
    Cavalcanti, A. N.
    ,
    M. M.Lobo
    ,
    C. M.Fontes
    ,
    P.Liporoni
    , and
    P.Mathias
    . 2005. Microleakage at the composite-repair interface: Effect of different surface treatment methods.Operative Dentistry30
    1
    :113117.
  • 7
    Burtscher, P.
    1993. Stability of radicals in cured composite materials.Dental Materials9
    4
    :218221.
  • 8
    Papacchini, F.
    ,
    S.Dall'Oca
    ,
    N.Chieffi
    ,
    C.Goracci
    ,
    F. T.Sadek
    ,
    B. I.Suh
    ,
    F. R.Tay
    , and
    M.Ferrari
    . 2007. Composite-to-composite microtensile bond strength in the repair of a micro-filled hybrid resin: Effect of surface treatment and oxygen inhibition.Journal of Adhesive Dentistry9
    1
    :2531.
  • 9
    Dall'Oca, S.
    ,
    F.Papacchini
    ,
    C.Goracci
    ,
    A. H.Cury
    ,
    B. I.Suh
    ,
    F. R.Tay
    , and
    M.Ferrari
    . 2006. Effect of oxygen inhibition on composite repair strength over time.Journal of Biomedical Material Research Part B: Applied Biomaterials81B:493498.
  • 10
    Brosh, T.
    ,
    R.Pilo
    ,
    N.Bichacho
    , and
    R.Blutstein
    . 1997. Effect of combinations of surface treatments and bonding agents on the bond strength of repaired composites.Journal of Prosthetic Dentistry77
    2
    :122126.
  • 11
    Matsumura, H.
    ,
    N.Hisamatsu
    , and
    M.Atsuta
    . 1995. Effect of unfilled resins and a silane primer on bonding between layers of a light-activated composite resin veneering material.Journal of Prosthetic Dentistry73
    4
    :386391.
  • 12
    Shahdad, S. A.
    and
    J. G.Kennedy
    . 1998. Bond strength of repaired anterior composite resins: An in vitro study.Journal of Dentistry26
    8
    :685694.
  • 13
    Bonstein, T.
    ,
    D.Garlapo
    ,
    J.DonarummoJr
    , and
    P. J.Bush
    . 2005. Evaluation of varied repair protocols applied to aged composite resin.Journal of Adhesive Dentistry7
    1
    :4149.
  • 14
    Hannig, C.
    ,
    S.Laubach
    ,
    P.Hahn
    , and
    T.Attin
    . 2006. Shear bond strength of repaired adhesive filling materials using different repair procedures.Journal of Adhesive Dentistry8
    1
    :3540.
  • 15
    Özcan, M.
    ,
    P.Alander
    ,
    P. K.Vallittu
    ,
    M. C.Huysmans
    , and
    W.Kalk
    . 2005. Effect of three surface conditioning methods to improve bond strength of particulate filler resin composites.Journal of Materials Science: Materials in Medicine16
    1
    :2127.
  • 16
    Yekta-Fard, M.
    and
    A. B.Ponter
    . 1992. Factors affecting the wettability of polymer surfaces.Journal of Adhesion Science and Technology6
    2
    :253277.
  • 17
    Teixeira, E. C.
    ,
    S. C.Bayne
    ,
    J. Y.Thompson
    ,
    A. V.Ritter
    , and
    E. J.Swift
    . 2005. Shear bond strength of self-etching bonding systems in combination with various composites used for repairing aged composites.Journal of Adhesive Dentistry7
    2
    :159164.
  • 18
    Papacchini, F.
    ,
    I.Radovic
    ,
    E.Magni
    ,
    C.Goracci
    ,
    F.Monticelli
    ,
    N.Chieffi
    ,
    A.Polimeni
    , and
    M.Ferrari
    . Flowable composites as intermediate agents without adhesive application in resin composite repair: A bond strength and SEM evaluation.American Journal of Dentistryin press.
  • 19
    Frankenberger, R.
    ,
    N.Krämer
    ,
    J.Ebert
    ,
    U.Lohbauer
    ,
    S.Käppel
    ,
    S.ten Weges
    , and
    A.Petschelt
    . 2003. Fatigue behaviour of the resin-resin bond of partially replaced resin-based composite restorations.American Journal of Dentistry16
    1
    :1722.
  • 20
    Frankenberger, R.
    ,
    S.Roth
    ,
    N.Krämer
    ,
    M.Pelka
    , and
    A.Petschelt
    . 2003. Effect of preparation mode on Class II resin composite repair.Journal of Oral Rehabilitation30
    6
    :559564.
  • 21
    Shen, C.
    ,
    E.Mondragon
    ,
    V. V.Gordan
    , and
    I. A.Mjör
    . 2004. The effect of mechanical undercuts on the strength of composite repair.Journal of the American Dental Association135
    10
    :14061412.
  • 22
    Daronch, M.
    ,
    F. A.Rueggeberg
    , and
    M. F.de Goes
    . 2005. Monomer conversion of pre-heated composite.Journal of Dental Research84
    7
    :663667.
  • 23
    Knight, J. S.
    ,
    R.Fraughn
    , and
    D.Norrington
    . 2006. Effect of temperature on the flow properties of resin composite.General Dentistry54
    1
    :1416.
  • 24
    Lee, J. H.
    ,
    C. M.Um
    , and
    I. B.Lee
    . 2006. Rheological properties of resin composites according to variations in monomer and filler composition.Dental Materials22
    6
    :515526.
  • 25
    Pashley, D. H.
    ,
    R. M.Carvalho
    ,
    H.Sano
    ,
    M.Nakajima
    ,
    M.Yoshiyama
    ,
    Y.Shono
    ,
    C. A.Fernandes
    , and
    F. R.Tay
    . 1999. The microtensile bond strength: A review.Journal of Adhesive Dentistry1
    4
    :299309.
  • 26
    Perdigão, J.
    ,
    S.Geraldeli
    ,
    A. R.Carmo
    , and
    H. R.Dutra
    . 2002. In vivo influence of residual moisture on microtensile bond strengths of one-bottle adhesives.Journal of Esthetic and Restorative Dentistry14
    1
    :3138.
  • 27
    Barnes, H. A.
    ,
    J. F.Hutton
    , and
    K.Walters
    . 1989. An Introduction to Rheology.
    Elsevier
    .
    New York
    .
  • 28
    Lee, I. B.
    ,
    H. H.Son
    , and
    C. M.Um
    . 2003. Rheologic properties of flow-able, conventional hybrid, and condensable composite resins.Dental Materials19
    4
    :298307.
  • 29
    Bayne, S. C.
    ,
    J. Y.Thompson
    ,
    E. J.SwiftJr
    ,
    P.Stamatiades
    , and
    M.Wilkerson
    . 1998. A characterization of first-generation flowable composites.Journal of the American Dental Association129
    5
    :567577.
  • 30
    Bonilla, E. D.
    ,
    M.Yashar
    , and
    A. A.Caputo
    . 2003. Fracture toughness of nine flowable resin composites.Journal of Prosthetic Dentistry89
    3
    :261267.
  • 31
    Clelland, N. L.
    ,
    M. P.Pagnotto
    ,
    R. E.Kerby
    , and
    R. R.Seghi
    . 2005. Relative wear of flowable and highly filled composite.Journal of Prosthetic Dentistry93
    2
    :153157.
  • 32
    Ferracane, J. L.
    2005. Developing a more complete understanding of stresses produced in dental composites during polymerization.Dental Materials21
    1
    :3642.
  • 33
    Kleverlaan, C. J.
    and
    A. J.Feilzer
    . 2005. Polymerization shrinkage and contraction stress of dental resin composites.Dental Materials21
    12
    :11501157.
  • 34
    Labella, R.
    ,
    P.Lambrechts
    ,
    B.Van Meerbeek
    , and
    G.Vanherle
    . 1999. Polymerization shrinkage and elasticity of flowable composites and filled adhesives.Dental Materials15
    2
    :128137.
  • 35
    Ausiello, P.
    ,
    A.Apicella
    , and
    C. L.Davidson
    . 2002. Effect of adhesive layer properties on stress distribution on composite restorations—a 3D finite element analysis.Dental Materials18
    4
    :295303.
  • 36
    Tjandrawinata, R.
    ,
    M.Irie
    , and
    K.Suzuki
    . 2005. Flexural properties of eight flowable light-cured restorative materials, in immediate vs 24-hour water storage.Operative Dentistry30
    2
    :239249.
  • 37
    Opdam, N. J.
    ,
    J. J.Roeters
    ,
    T.de Boer
    ,
    D.Pesschier
    , and
    E.Bronkhorst
    . 2003. Voids and porosities in Class I micropreparations filled with various resin composites.Operative Dentistry28
    1
    :914.
  • 38
    Ölmez, A.
    ,
    N.Öztas
    , and
    H.Bodur
    . 2004. The effect of flowable resin composite on microleakage and internal voids in Class II composite restorations.Operative Dentistry29
    6
    :713719.
  • 39
    Li, Q.
    ,
    S.Jepsen
    ,
    H. K.Albers
    , and
    J.Eberhard
    . 2006. Flowable materials as an intermediate layer could improve the marginal and internal adaptation of composite restorations in Class-V-cavities.Dental Materials22
    3
    :250257.
  • 40
    Trujillo, M.
    ,
    S. M.Newman
    , and
    J. W.Stansbury
    . 2004. Use of near-IR to monitor the influence of external heating on dental composite photopolymerization.Dental Materials20
    8
    :766777.
  • 41
    Daronch, M.
    ,
    F. A.Rueggeberg
    ,
    M. F.de Goes
    , and
    R.Giudici
    . 2006. Polymerization kinetics of pre-heated composite.Journal of Dental Research85
    1
    :3843.
  • 42
    Tiba, A.
    ,
    D. G.Charlton
    ,
    K. S.Vandewalle
    , and
    M. E.Cohen
    . 2005. Volumetric polymerization shrinkage of resin composites under simulated intraoral temperature and humidity conditions.Operative Dentistry30
    6
    :696701.
  • 43
    Plasmans, P. J.
    ,
    N. H.Creugers
    ,
    R. J.Hermsen
    , and
    M. M.Vrijhoef
    . 1994. Intraoral humidity during operative procedures.Journal of Dentistry22
    2
    :8991.
  • 44
    Friedman, J.
    2001. Method for pre-heating dental materials.United States Patent, 6 320 162.
  • 45
    Gregory, W. A.
    ,
    B.Pounder
    , and
    E.Bakus
    . 1990. Bond strengths of chemically dissimilar repaired composite resins.Journal of Prosthetic Dentistry64
    6
    :664668.
  • 46
    Papacchini, F.
    ,
    F. L.de Castro
    ,
    C.Goracci
    ,
    T. N.Sardella
    ,
    F. R.Tay
    ,
    A.Polimeni
    ,
    M.Ferrari
    , and
    R. M.Carvalho
    . 2005. An investigation of the contribution of silane to the composite repair strength over time using a double-sided microtensile test.International Dentistry South Africa8
    1
    :2636.
Copyright: © 2007 This is an Open Access article: verbatim copying and redistribution of this article are permitted in all media for any purpose, provided this notice is preserved along with the article's original URL. 2007
Figure 1
Figure 1

Distribution of failure modes for all experimental groups. Values are the percentage of sticks fractured in the respective failure mode within each subgroup.


Figure 2
Figure 2

Representative Gradia-to-Gradia interfaces repaired by direct bonding of the repairing composite (RC) to the sandblasted composite substrate (CS) (bar=50 μm). Figure 2A. SEM micrograph of CS repaired by using 4°C-cooled RC. Gaps (20–100 μm) and discontinuities interrupt the interfacial coupling. Figure 2B. SEM micrograph of CS repaired with 37°C pre-heated RC. A good interfacial composite-to-composite adaptation barely interrupted by small defect (less than 15 μm) is detectable.


Figure 2
Figure 2

Continued.


Figure 3
Figure 3

Representative Gradia-to-Gradia interfaces repaired by unfilled resin Scotchbond Multi-Purpose Adhesive as an intermediate agent (RC: repairing composite adhesive resin, CS: composite substrate) (bar=50 μm). Figure 3A: Voids and defects (10–70 μm) both in the RC and in close proximity of the adhesive layer (A) are detectable with 4°C-RC. Figure 3B. A uniform adaptation at the CS-A and A-RC interfaces are evident with 23°C-RC.


Figure 3
Figure 3

Continued.


Figure 4
Figure 4

Representative Gradia-to-Gradia interfaces repaired by flowable composite Gradia LoFlo as an intermediate agent (RC: repairing composite, FC: flowable composite, CS: composite substrate) (bar=50 μm). Fiigure 4A. Particular of 4°C RC-FC interface. A gap (50–100 μm), probably resulting from difficult adaptation of RC is clearly shown. Figure 4B. 23°C-composite as RC. Good coupling at both CS-FC and FC-RC interfaces is visible.


Figure 4
Figure 4

Continued.


Contributor Notes

Federica Papacchini, DDS, MSc, PhD, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

Elisa Magni, DDS, MSc, PhD student, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

Ivana Radovic, DDS, MSc, clinical assistant, Clinic for Pediatric and Preventive Dentistry, Faculty of Dentistry, University of Belgrade, Serbia; PhD student, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

Claudia Mazzitelli, DDS, MSc, PhD student, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

Francesca Monticelli, DDS, MSc, PhD, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy; post-doctoral researcher, Department of Dental Materials, School of Dentistry, University of Granada, Spain

Cecilia Goracci, DDS, MSc, PhD, assistant professor, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

Antonella Polimeni, DMD, DDS, professor, Department of Pediatric Dentistry, University “La Sapienza,” Rome, Italy

Marco Ferrari, DMD, DDS, PhD, professor, Department of Dental Materials and Restorative Dentistry, University of Siena, Siena, Italy

*Reprint request: Policlinico Le Scotte, Viale Bracci, Siena, PO Box 53100, Italy; e-mail: md3972@mclink.it
Received: 11 Aug 2006
  • Download PDF