Controlling In Vivo, Human Pulp Temperature Rise Caused by LED Curing Light Exposure

,
,
,
,
, and
Page Range: 235 – 241
DOI: 10.2341/17-364-C
Save
Download PDF

SUMMARY

Objective:

The objective of this study was to evaluate the in vivo effectiveness of air spray to reduce pulp temperature rise during exposure of intact premolars to light emitted by a high-power LED light-curing unit (LCU).

Methods and Materials:

After local Ethics Committee approval (#255945), intact, upper first premolars requiring extraction for orthodontic reasons from five volunteers received infiltrative and intraligamental anesthesia. The teeth (n=9) were isolated using rubber dam, and a minute pulp exposure was attained. The sterile probe from a wireless, NIST-traceable, temperature acquisition system was inserted directly into the coronal pulp chamber. Real-time pulp temperature (PT) (°C) was continuously monitored, while the buccal surface was exposed to a polywave LED LCU (Bluephase 20i, Ivoclar Vivadent) for 30 seconds with simultaneous application of a lingually directed air spray (30s-H/AIR) or without (30s-H), with a seven-minute span between each exposure. Peak PT values were subjected to one-way, repeated-measures analysis of variance, and PT change from baseline (ΔT) during exposure was subjected to paired Student's t-test (α=0.05).

Results:

Peak PT values of the 30s-H group were significantly higher than those of 30s-H/AIR group and those from baseline temperature (p<0.001), whereas peak PT values in the 30s-H/AIR group were significantly lower than the baseline temperature (p=0.003). The 30s-H/AIR group showed significantly lower ΔT values than did the 30s-H group (p<0.001).

Conclusion:

Applying air flow simultaneously with LED exposure prevents in vivo pulp temperature rise.

INTRODUCTION

Light-emitting diode (LED) light-curing units (LCUs) have become part of everyday clinical dentistry. The success of these devices is due to their efficiency and design, as they are battery driven and often do not require cooling fans. In addition, these LCUs are able to emit light with considerably higher radiant emittance values and have higher source longevity than do conventional quartz-tungsten-halogen lights (QTH).

Over the last decade, powerful LED LCUs have become commercially available to promote better polymerization of resin-based materials using shorter exposure durations. These devices are capable of emitting light with irradiance over 2000 mW/cm2, and exposure times of no longer than five seconds are recommended by some manufacturers. As a consequence, these high-power LED LCUs generate heat in the target object.1 In this regard, several in vitro studies report a significant temperature increase within the pulp chamber of extracted teeth, ranging from 1.5°C to 23.2°C.2-8 Although this temperature range depends on LCU type, radiant emittance, and tooth characteristics,3-12 there is a consensus that the use of some LED LCUs can result in a pulp temperature (PT) rise to values close to, or even higher than, the threshold temperature increase of 5.5°C, a value considered harmful to the pulp.13 Most recently, a new in vivo methodology comprising the placement of a temperature probe within the pulp in human premolars has been developed.14,15 Using this method, higher PT rise than 5.5 °C in intact premolars was observed as radiant exposure value increased.15

Heat control during restorative procedures to assure pulp integrity has been a great concern among clinicians and researchers.13,16,17 For instance, several studies evaluated the effectiveness of alternative methods to reduce the heat caused during tooth preparation when using slow- and high-speed hand pieces and have shown it to be an effective method in reducing heat generated during that restorative step.17-22 Other studies evaluated the use of different lasers instead of burs,23-27 as well as the use of different cooling system protocols, including air and water spray from the triple syringe.18,28-30 In addition, some cooling strategies have proven to be effective in reducing temperature rise within the pulp chamber during the debonding of orthodontic brackets31 or during the fabrication of provisional restorations comprising the use of self-polymerized acrylic resins.32 In this regard, different approaches have been suggested to reduce heat during the polymerization of resin composites, such as the use of reduced irradiance,30,33 or discontinuous curing modes.34 However, such procedures may compromise monomer conversion and impair the mechanical properties of the resulting resin composites.35 To date, only one in vitro study demonstrated that cooling strategies comprising the use of air or the combination of air/water can reduce temperature rise during exposure of indirect restorations to curing light.36 Another study showed that a cooling system installed in a LED LCU was effective in reducing the maximum temperature rise within the pulp chamber and speeding up the cooling process of the tooth after light shut off.4 Based on these findings, it is reasonable to assume that the use of air spray would prevent PT rise in vivo, in human premolars, during exposure to the light emitted from a polywave LED LCU (Bluephase 20i, Ivoclar Vivadent, Schann, Linchestein Principality). However, no in vivo study has evaluated the effectiveness of such procedures during exposure to a curing light.

Thus, the purpose of this in vivo study was to measure and evaluate PT changes when intact human premolars were exposed to a high-intensity LED LCU, while either simultaneously applying an air spray from the opposite side of the exposure or not. The research hypotheses were that 1) the application of air spray during LCU exposure will significantly reduce PT rise in comparison to when no air spray is used, and 2) PT values following light exposure with air spray will result in PT values significantly lower than pre-exposure, baseline values.

METHODS AND MATERIALS

Irradiance Measurement of the Polywave LED LCU

The spectral power of the polywave LED light was recorded five times in High mode, using a laboratory grade spectroradiometer (USB 2000, Ocean Optics, Dunedin, FL, USA) and a 6-in integrating sphere (Labsphere, North Sutton, NH, USA), previously calibrated using a National Institute of Standards and Technology (NIST)-traceable light source. The LCU tip end was positioned at the entrance of the integrating sphere, so all light emitted from the unit was captured. Spectral power was measured between 350 and 550 nm, using software (Spectra-Suite v2.0.146, Ocean Optics), which also provided a total value within that range. The optical emitting area of the distal end of the light guide was calculated, and this value was divided into the integrated spectral power value to derive the total radiant exitance from the curing light.

In Vivo Measurement of Intrapulpal Temperature

The study was previously approved by the local Ethics Committee (protocol #255945). Five patients requiring extraction of first premolars for orthodontic reasons (n=9) were recruited from the orthodontic specialization programs in Ponta Grossa, State University of Ponta Grossa and the Brazilian Association of Dentistry, Parana section. The subjects, four females and one male, ranging from 12 to 30 years old, passed through an initial consultation phase, in which the volunteers were informed about the study aims and all methodology involved in the research. Patient inclusion criteria included 1) treatment plans indicating premolar extractions for orthodontic reasons, 2) the presence of healthy, intact, noncarious, and nonrestored, fully erupted treatment teeth, and 3) patients with well-controlled health conditions that allowed all procedures involved in the research to be performed with minimal risk. Exclusion criteria included those patients who were currently under medication. After having the benefits and risks for participating in the study explained, the volunteer subjects, or the subject's legally authorized representative (for subjects under the age of 18 years), signed the informed consent document. The subject's medical history was obtained and reviewed by at least one clinician participating in the study, prior to any treatment being performed.

All teeth tested were anesthetized using a local anesthetic (2% mepivacaine hydrochloride with epinephrine, Mepiadre, DFL Industria e Comercio, Rio de Janeiro, RJ, Brazil). Only intact, upper first premolars were evaluated. The teeth received both infiltrative as well as intraligamental injections. The premolar received isolation with rubber dam. A deep class I preparation was then made on the occlusal surface, using a round diamond bur (#1013, KG Sorensen, Cotia, SP, Brazil), under constant irrigation, until the preparation floor was near the roof of the pulp chamber. The pulp exposure was carefully obtained with the aid of a root canal explorer, so the exposure diameter was as small as possible. After pulp exposure was obtained, a sterile thermocouple probe from a wireless temperature acquisition system (Temperature Data Acquisition-Thermes Wfi, Physitemp, Clifton, NJ, USA) was inserted directly into the pulp chamber. The thermocouple probe was connected to an isolated, battery-driven data acquisition device that wirelessly sent temperature values to a computer, where temperature was displayed and recorded in real time. With the probe remaining in place, the occlusal preparation was then filled using a provisional restorative material (Cavitec, CaiTHEC Ltda, São José dos Pinhais, PR, Brazil), to minimize heat loss from the tooth through the cavity walls and pulp access. The baseline pulpal temperature value was established after approximately 15 to 22 minutes. The emitting end of a high power, polywave LED curing unit (Bluephase 20i, Ivoclar Vivadent), used in its high-power mode (1244 mW/cm2), was placed directly against the facial enamel surface, in the cervical region. The following exposure conditions were then applied, during which data were recorded at every 0.2 seconds: three seconds prior to light exposure, an air spray from a triple syringe was held 1-mm from the lingual enamel surface, opposite of where the LCU tip was placed (air pressure set to 28 psi). A 30-second exposure in high-intensity mode (radiant exposure of 37.3 J/cm2) was applied to the facial surface, and the air stream was terminated at the same moment the LCU turned off (30s-H/AIR). A seven-minute time span then passed, to allow baseline pulpal temperatures to return to the pre-exposure values. Another exposure was then given as stated before, but without application of the air spray (30s-H). At the end of temperature acquisition, the probe was carefully removed from the tooth. After testing was completed, the teeth were atraumatically extracted, as planned for the patient's orthodontic treatment. Following extraction, the probe was reinserted into the pulp chamber of some extracted teeth, and X-rays were taken from the proximal surface, with the probe in position as it was intraorally, to verify probe position during testing (Figure 1).

Figure 1. X-ray analysis after tooth extraction to confirm if the probe was properly positioned into the pulp chamber during data acquisition.Figure 1. X-ray analysis after tooth extraction to confirm if the probe was properly positioned into the pulp chamber during data acquisition.Figure 1. X-ray analysis after tooth extraction to confirm if the probe was properly positioned into the pulp chamber during data acquisition.
Figure 1 X-ray analysis after tooth extraction to confirm if the probe was properly positioned into the pulp chamber during data acquisition.

Citation: Operative Dentistry 44, 3; 10.2341/17-364-C

Statistical Analyses

Peak PT during exposure and PT increase during exposure relative to the pre-exposure baseline value (ΔT) were determined. Peak PT data were evaluated using a one-way, repeated-measures analysis of variance, whereas ΔT was compared using a paired Student's t-test. All statistical testing was made using a preset α of 0.05. Post hoc power analysis was performed for statistical analyses of peak PT and ΔT values. All analyses were performed using commercial statistical software (Statistics 19, SPSS Inc, IBM Company, Armonk, NY, USA).

RESULTS

In Vivo Measurement of Intrapulpal Temperature

For the number of evaluated teeth (n=9), the in vivo study was adequately powered for both peak PT and ΔT values (>99.0%; α=0.05). The results of peak PT and ΔT are shown in Table 1. The PT from exposure to the 30s-H condition (37.5±0.5°C) was significantly greater than that of the baseline PT (35.4±0.6°C). On the other hand, the use of the 30s-H/AIR condition resulted in significantly lower PT values (34.0±1.0°C) than the baseline (35.4±0.6°C) (p=0.003). As a result, the ΔT values observed when teeth were exposed to a 30s-H mode were significantly higher (p<0.001) than those observed using the 30s-H/AIR condition.

Table 1 Means (SD) of Peak Temperature Values (°C) and ΔT (°C) for Each Experimental Conditiona
Table 1

Figure 2 shows representative real-time profiles of PT change during exposure to 30s-H and 30s-H/AIR conditions. Exposure to the 30s-H mode resulted in an immediate PT increase, characterized by an ascending straight line, which was followed by a slow temperature drop, following light termination. On the other hand, exposure to the 30s-H/AIR condition caused a slow PT decrease while the buccal surface was exposed to curing light. When the light terminated and the air spray stopped, PT dropped more quickly, for approximately 30 seconds, before it started to rise again.

Figure 2. Representative, in vivo, real-time PT profiles during a 30-s exposure on high output mode (30s-H) without (blue line) or with (30s-H/AIR) (orange line) simultaneous application of air spray from the opposite tooth side. Duration of air-spray depicted by the light yellow rectangle, and duration of light exposure represented by the blue box.Figure 2. Representative, in vivo, real-time PT profiles during a 30-s exposure on high output mode (30s-H) without (blue line) or with (30s-H/AIR) (orange line) simultaneous application of air spray from the opposite tooth side. Duration of air-spray depicted by the light yellow rectangle, and duration of light exposure represented by the blue box.Figure 2. Representative, in vivo, real-time PT profiles during a 30-s exposure on high output mode (30s-H) without (blue line) or with (30s-H/AIR) (orange line) simultaneous application of air spray from the opposite tooth side. Duration of air-spray depicted by the light yellow rectangle, and duration of light exposure represented by the blue box.
Figure 2 Representative, in vivo, real-time PT profiles during a 30-s exposure on high output mode (30s-H) without (blue line) or with (30s-H/AIR) (orange line) simultaneous application of air spray from the opposite tooth side. Duration of air-spray depicted by the light yellow rectangle, and duration of light exposure represented by the blue box.

Citation: Operative Dentistry 44, 3; 10.2341/17-364-C

DISCUSSION

The data clearly indicate that air spray application during light exposure was able to prevent PT to rise. Therefore, the first research hypothesis was not rejected. These results are in agreement with those from a previous in vitro study,36 which attributed the ability of air flow to remove the hot air around an exposed tooth. According to the authors, the hot air around the exposed tooth would be partly responsible for the temperature increase within the tooth. Moreover, the current results can also be attributed to the thermal behavior of enamel and dentin. In this regard, when an intact tooth is exposed to a curing light, the thermal energy from enamel dissipates inward, toward the dentin and pulp chamber, as well as outward to the ambient environment.37 Because dentin has a low thermal diffusivity, the thermal energy is not only transferred to the pulp but is also stored within it.37 This thermal behavior is responsible for gradual dissipation of heat energy toward the pulp, sustaining higher temperature values within the pulp, even after the curing light has shut off, as previously shown.15 Conversely, when an air flow was started three seconds before the light was turned on and was maintained during light exposure, the temperature of enamel and dentin dropped. Thus, higher thermal energy was required to heat these substrates. Therefore, when air flow and exposure to light occurred simultaneously, the heat generated during the 30-second exposure to the curing light emitting 1244 mW/cm2 (approximately 37.3 J/cm2) was not capable of increasing the temperature of enamel and dentin to values that could cause a PT rise. Indeed, based on the real-time analysis of temperature change during a 30-second exposure to light with air flow, a slow PT decrease was noted during exposure to curing light (Figure 2). This finding infers that the use of air flow during exposure to a curing light was more effective in decreasing PT than was light in transferring heat to the pulp. For this reason, it is reasonable to assume that longer exposures to a curing light emitting 1244 mW/cm2, which is an irradiance value commonly found in commercially available LED LCUs, would not be harmful for the pulp when air flow is applied simultaneously.

Interestingly, the temperature drop in dentin due to air flow not only avoided PT rise during exposure to the LED light but also decreased PT to lower values than that of the baseline PT (Table 1). As a matter of fact, PT values showed a continuing fast decrease for approximately 30 seconds after air flow stopped (Figure 2). Therefore, the second hypothesis was not rejected. This finding confirms the presence of a gradient temperature created between the pulp and the cooled pulp chamber walls, due to the decrease in dentin temperature during air flow. In other words, the lower temperature of dentin chamber walls, in comparison to that of the pulp, causes the heat from the pulp to flow toward the dentin chamber walls by thermal conduction, to reach thermodynamic equilibrium, resulting in the continuous post-air flow PT drop. In a restorative procedure with photo-activated resin-based composites, when sequential tooth exposures to the LED light are required to ensure optimal polymerization of resin composite layers, a prepared tooth having PT lower than the baseline temperature would provide further protection against pulp damage. In that clinical condition, higher radiant exposure values would be required to increase PT to the threshold values considered harmful for the pulp (approximately 42.5°C)13 than when such restorative procedures are performed on teeth having PT at the baseline values. Further investigation is required to confirm such assumptions.

The current results demonstrate that air flow was very effective in preventing PT rise during exposure to an LED light using its “high” mode in intact premolars in vivo. In this context, it is worth noticing that PT rise may be influenced by the translucency and thermal diffusivity of the substrate interposed between the heat source and the pulp.15,38 Because resin composites may show a wide range in translucency39 and lower thermal diffusivity depending on the amount and type of filler content,40 PT rise may be higher during the exposure of a class V preparation to curing light when the prepared cavity is filled with more translucent resin composites having low filler content in comparison to that observed on intact teeth. For instance, previous in vitro findings have shown that exposing teeth with a class V preparation either empty or filled with resin composite to a curing light caused higher PT rise than did exposure of an intact tooth to a curing light.41 As a consequence, one could state that air flow may not be as effective in these clinical scenarios as it was in intact teeth. For this reason, in an attempt to overcome this study limitation, the intact teeth were exposed to LED light for 30 seconds, which is longer than 10- or 20-second exposure usually recommended by the manufacturers of most recently commercially available resin composites. In other words, because of the direct positive relationship between exposure interval and PT rise,15 the 30-second exposure to light may cause higher PT rise than that expected even when a class V preparation filled with translucent resin composites with high thermal diffusivity is exposed to shorter exposure intervals, such as 10- or 20-second exposure to LED light. For this reason, it is reasonable to assume that applying air flow along with the curing light exposure in these restorative procedures may be as effective as the air flow applied in the condition evaluated in the current study. In addition, because intact teeth were evaluated, similar results could also be expected in procedures involving the delivery of curing light to intact teeth, such as in-office bleaching, when cementing ceramic laminates, or when placing composite resin veneers.1,42

Based on these findings, exposure of intact premolars to a curing light along with simultaneous application of an air flow is shown to be an effective method to prevent temperature rise within the pulp during exposure to high irradiance levels. However, because of differences in morphology and dentin thickness among teeth, care should be taken when assuming that similar results should be expected on other teeth, such as anterior incisors. Despite this limitation, such an easy and time-efficient approach should be added to the clinician's routine, because it can reduce the risk of pulp thermal damage when powerful LED devices are used.

CONCLUSION

Within the limitations imposed by this in vivo study, it is possible to conclude that 1) air flow applied simultaneously with exposure to LED light is capable of preventing a temperature increase within the pulp during curing light exposure. In addition, 2) this procedure causes a further temperature drop in the pulp after exposure to a curing light.

Acknowledgments

The authors are indebted to State University of Ponta Grossa and Augusta University for all support required to perform this study. The authors thank Ivoclar Vivadent for the donation of the light curing unit used in the current study. This study was supported by grants from the Araucaria Foundation (#232/2014 and #488/2014).

Regulatory Statement

This study was conducted in accordance with all the provisions of the local human subjects oversight committee guidelines and policies of approval of the State University of Ponta Grossa. The approval code for this study is #255945.

Conflict of Interest

The authors of this manuscript certify that they have no proprietary, financial, or other personal interest of any nature or kind in any product, service, and/or company that is presented in this article.

REFERENCES

  • 1
    Asmussen E,
    &
    Peutzfeldt A
    (2005) Temperature rise induced by some light emitting diode and quartz-tungsten-halogen curing unitsEuropean Journal of Oral Science113(
    1
    ) 96-98.
  • 2
    Kodonas K,
    Gogos C,
    &
    Tziafa C
    (2009) Effect of simulated pulpal microcirculation on intrachamber temperature changes following application of various curing units on tooth surfaceJournal of Dentistry37(
    6
    ) 485-490.
  • 3
    Baroudi K,
    Silikas N,
    &
    Watts DC
    (2009) In vitro pulp chamber temperature rise from irradiation and exotherm of flowable compositesInternational Journal of Paediatric Dentistry19(
    1
    ) 48-54.
  • 4
    Park SH,
    Roulet JF,
    &
    Heintze SD
    (2010) Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow ratesOperative Dentistry35(
    3
    ) 353-361.
  • 5
    Yazici AR,
    Muftu A,
    Kugel G,
    &
    Perry RD
    (2006) Comparison of temperature changes in the pulp chamber induced by various light curing units, in vitroOperative Dentistry31(
    2
    ) 261-265.
  • 6
    Eldeniz AU,
    Usumez A,
    Usumez S,
    &
    Ozturk N
    (2005) Pulpal temperature rise during light-activated bleachingJournal of Biomedical Materials Research B Applied Biomaterials72(
    2
    ) 254-259.
  • 7
    Leprince J,
    Devaux J,
    Mullier T,
    Vreven J,
    &
    Leloup G
    (2010) Pulpal-temperature rise and polymerization efficiency of LED curing lightsOperative Dentistry35(
    2
    ) 220-230.
  • 8
    Oberholzer TG,
    Makofane ME,
    du Preez IC,
    &
    George R
    (2012) Modern high powered led curing lights and their effect on pulp chamber temperature of bulk and incrementally cured composite resinEuropean Journal of Prosthodontics and Restorative Dentistry20(
    2
    ) 50-55.
  • 9
    Millen C,
    Ormond M,
    Richardson G,
    Santini A,
    Miletic V,
    &
    Kew P
    (2007) A study of temperature rise in the pulp chamber during composite polymerization with different light-curing unitsJournal of Contemporary Dental Practice8(
    7
    ) 29-37.
  • 10
    Baik JW,
    Rueggeberg FA,
    &
    Liewehr FR
    (2001) Effect of light-enhanced bleaching on in vitro surface and intrapulpal temperature riseJournal of Esthetic and Restorative Dentistry13(
    6
    ) 370-378.
  • 11
    Gomes M,
    DeVito-Moraes A,
    Francci C,
    Moraes R,
    Pereira T,
    Froes-Salgado N,
    Yamazaki L,
    Silva L,
    &
    Zezell D
    (2013) Temperature increase at the light guide tip of 15 contemporary LED units and thermal variation at the pulpal floor of cavities: an infrared thermographic analysisOperative Dentistry38(
    3
    ) 324-333.
  • 12
    Rajesh Ebenezar AV,
    Anilkumar R,
    Indira R,
    Ramachandran S,
    &
    Srinivasan MR
    (2010) Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro studyJournal of Conservative Dentistry13(
    3
    ) 132-135.
  • 13
    Zach L,
    &
    Cohen G
    (1965) Pulp response to externally applied heatOral Surgery, Oral Medicine and Oral Pathol19(
    4
    ) 515-530.
  • 14
    Runnacles P,
    Arrais CA,
    Pochapski MT,
    dos Santos FA,
    Coelho U,
    Gomes JC,
    De Goes MF,
    Gomes OM,
    &
    Rueggeberg FA
    (2015) Direct measurement of time-dependent anesthetized in vivo human pulp temperatureDental Materials31(
    1
    ) 53-59.
  • 15
    Runnacles P,
    Arrais CA,
    Pochapski MT,
    Dos Santos FA,
    Coelho U,
    Gomes JC,
    De Goes MF,
    Gomes OM,
    &
    Rueggeberg FA
    (2015) In vivo temperature rise in anesthetized human pulp during exposure to a polywave LED light curing unitDental Materials31(
    5
    ) 505-513.
  • 16
    Schuchard A,
    &
    Watkins CE
    (1965) Thermal and histologic response to high-speed and ultrahigh-speed cutting in tooth structureJournal of American of Dental Association71(
    6
    ) 1451-1458.
  • 17
    Kwon SJ,
    Park YJ,
    Jun SH,
    Ahn JS,
    Lee IB,
    Cho BH,
    Son HH,
    &
    Seo DG
    (2013) Thermal irritation of teeth during dental treatment proceduresRestorative Dentistry, & Endodontics38(
    3
    ) 105-112.
  • 18
    Cavalcanti BN,
    Otani C,
    &
    Rode SM
    (2002) High-speed cavity preparation techniques with different water flowsJournal of Prosthetic Dentistry87(
    2
    ) 158-161.
  • 19
    Stanley HR, Jr.,
    &
    Swerdlow H
    (1959) Reaction of the human pulp to cavity preparation: results produced by eight different operative grinding technicsJournal of American Dental Association58(
    5
    ) 49-59.
  • 20
    Terranova PL
    (1967) Adverse conditions found in the use of ultra high-speed equipmentNew York State Dental Journal33(
    3
    ) 143-148.
  • 21
    Bhandary N,
    Desai A,
    &
    Shetty YB
    (2014) High speed handpiecesJournal of International Oral Health6(
    1
    ) 130-132.
  • 22
    Peyton FA
    (1955) Temperature rise in teeth developed by rotating instrumentsJournal of American Dental Association50(
    6
    ) 629-632.
  • 23
    Penn C,
    Beninati C,
    Mariano A,
    Dooley D,
    Harsono M,
    Perry R,
    &
    Kugel G
    (2014) Thermal effects on pulp due to laser and handpiece usageCompendium of Continuing Education in Dentistry35(
    10
    ) e41-e44.
  • 24
    Mollica FB,
    Camargo FP,
    Zamboni SC,
    Pereira SM,
    Teixeira SC,
    &
    Nogueira L, Jr.
    (2008) Pulpal temperature increase with high-speed handpiece, Er:YAG laser and ultrasound tipsJournal of Applied Oral Science16(
    3
    ) 209-213.
  • 25
    Krmek SJ,
    Miletic I,
    Simeon P,
    Mehicic GP,
    Anic I,
    &
    Radisic B
    (2009) The temperature changes in the pulp chamber during cavity preparation with the Er:YAG laser using a very short pulsePhotomedicine and Laser Surgery27(
    2
    ) 351-355.
  • 26
    Armengol V,
    Jean A,
    &
    Marion D
    (2000) Temperature rise during Er:YAG and Nd:YAP laser ablation of dentinJournal of Endodontics26(
    3
    ) 138-141.
  • 27
    Colucci V,
    do Amaral FL,
    Pecora JD,
    Palma-Dibb RG,
    &
    Corona SA
    (2009) Water flow on erbium:yttrium-aluminum-garnet laser irradiation: effects on dental tissuesLasers in Medical Science24(
    5
    ) 811-818.
  • 28
    Leung BT,
    Dyson JE,
    &
    Darvell BW
    (2012) Coolant effectiveness in dental cutting with air-turbine handpiecesNew Zealand Dentistry Journal108(
    1
    ) 25-29.
  • 29
    Henning G,
    &
    Przetak CC
    (1975) [The prep-trainer--thermal control in the pulp space]Deutsche Zahnarztliche Zeitschrift30(
    3
    ) 204-206.
  • 30
    Ozturk B,
    Usumez A,
    Ozturk AN,
    &
    Ozer F
    (2004) In vitro assessment of temperature change in the pulp chamber during cavity preparationJournal of Prosthetic Dentistry91(
    5
    ) 436-440.
  • 31
    Kley P,
    Frentzen M,
    Kupper K,
    Braun A,
    Kecsmar S,
    Jager A,
    &
    Wolf M
    (2016) Thermotransduction and heat stress in dental structures during orthodontic debonding: effectiveness of various cooling strategiesJournal of Orofacial Orthopedics77(
    3
    ) 185-193.
  • 32
    Moulding MB,
    &
    Loney RW
    (1991) The effect of cooling techniques on intrapulpal temperature during direct fabrication of provisional restorationsInternational Journal of Prosthodontics4(
    4
    ) 332-336.
  • 33
    Uhl A,
    Mills RW,
    &
    Jandt KD
    (2003) Polymerization and light-induced heat of dental composites cured with LED and halogen technologyBiomaterials24(
    10
    ) 1809-1820.
  • 34
    Hofmann N,
    Markert T,
    Hugo B,
    &
    Klaiber B
    (2003) Effect of high intensity vs. soft-start halogen irradiation on light-cured resin-based composites. Part I. Temperature rise and polymerization shrinkageAmerican Journal of Dentistry16(
    6
    ) 421-430.
  • 35
    Yap AU,
    Soh MS,
    &
    Siow KS
    (2002) Effectiveness of composite cure with pulse activation and soft-start polymerizationOperative Dentistry27(
    1
    ) 44-49.
  • 36
    Onisor I,
    Asmussen E,
    &
    Krejci I
    (2011) Temperature rise during photo-polymerization for onlay lutingAmerican Journal of Dentistry24(
    4
    ) 250-256.
  • 37
    Chiang YC,
    Lee BS,
    Wang YL,
    Cheng YA,
    Chen YL,
    Shiau JS,
    Wang DM,
    &
    Lin CP
    (2008) Microstructural changes of enamel, dentin-enamel junction, and dentin induced by irradiating outer enamel surfaces with CO2 laserLasers in Medical Science23(
    1
    ) 41-48.
  • 38
    Panas AJ,
    Zmuda S,
    Terpilowski J,
    &
    Preiskorn M
    (2003) Investigation of the thermal diffusivity of human tooth hard tissueInternational Journal of Thermophysics24(
    3
    ) 837-848.
  • 39
    Ryan EA,
    Tam LE,
    &
    McComb D
    (2010) Comparative translucency of esthetic composite resin restorative materialsJournal of the Canadian Dental Association76a84.
  • 40
    Watts DC,
    McAndrew R,
    &
    Lloyd CH
    (1987) Thermal diffusivity of composite restorative materialsJournal of Dental Research66(
    10
    ) 1576-1578.
  • 41
    Choi SH,
    Roulet JF,
    Heintze SD,
    &
    Park SH
    (2014) Influence of cavity preparation, light-curing units, and composite filling on intrapulpal temperature increase in an in vitro tooth modelOperative Dentistry39(
    5
    ) E195-E205.
  • 42
    Santini A,
    Watterson C,
    &
    Miletic V
    (2008) Temperature rise within the pulp chamber during composite resin polymerisation using three different light sourcesOpen Dentistry Journal2137-141.
Copyright: Operative Dentistry, 2019 2019
Figure 1
Figure 1

X-ray analysis after tooth extraction to confirm if the probe was properly positioned into the pulp chamber during data acquisition.


Figure 2
Figure 2

Representative, in vivo, real-time PT profiles during a 30-s exposure on high output mode (30s-H) without (blue line) or with (30s-H/AIR) (orange line) simultaneous application of air spray from the opposite tooth side. Duration of air-spray depicted by the light yellow rectangle, and duration of light exposure represented by the blue box.


Contributor Notes

Corresponding author: Av. Carlos Cavalcanti, 4748, Uvaranas, Ponta Grossa, Parana 08030-900, Brazil; e-mail: cesararrais@yahoo.com.br
Accepted: 02 Apr 2018
  • Download PDF